ACOUSTICS

a1aAcoustic Laboratory has developed high accuracy capability to support national acoustics measurement activities. The chart illustrates the traceability for acoustics measurement in Malaysia.

Slide1aaa

a1c

      1-inch and ½-inch laboratory standard microphone

The highest standards are 1-inch laboratory standard microphone (LS1P) and ½-inch laboratory standard microphone (LS2P) that are maintained using Reciprocity Calibration System. The frequency range is from 20 Hz to 25 kHz with the best uncertainty of 0.04 dB at 1 kHz. 

 

HIGH VOLTAGE

The high voltage and high current capabilities in the laboratory are described as follows:

 

Systems Parameter Capability Range Best Measurement Uncertainty (±)
Impulse high voltage Peak Voltage 50 kV - 500 kV 1.1%
Time 3.90%
AC high voltage AC high voltage 5 kV - 200 kV 0.66%
DC high voltage DC high voltage 5 kV - 150 kV 0.10%
> 150 kV - 270 kV 0.60%
Current Transformer Ratio Error Primary current: 5 A - 2000 A; Secondary current: 5 A 0.05%
Phase displacement 0.8'
Voltage Transformer Ratio Error Primary voltage: 3.3 kV - 11 kV; Secondary voltage: 110 V 0.05%
Phase displacement 0.8'

 

hv1hv2
hv3

AC & LOW FREQUENCY

AC & Low Frequency Laboratory are capable to generate and measure DC to 10 MHz and beyond covering voltage, current, and resistance parameters to support national AC & Low Frequency measurement activities. The chart illustrates the traceability for AC & Low Frequency in Malaysia.

ac1

The highest standards are primary Thermal Converters and AC DC Shunts maintained by utilizing comparison method of AC-DC. Different comparison methods for accurate RMS value determination of AC Voltage and AC Current at low frequencies are traceable to the national standard of AC Voltage and Current. The frequency range is from 10 Hz to 1 MHz with best uncertainty of 5 uV/V. 

ac1ac2

Thermal Converter

Reference Multimeter

MAGNETICS

m1bMagnetics Laboratory has developed high accuracy capability to support national magnetics measurement activities. The chart illustrates the traceability for magnetics measurement in Malaysia.

 m1a

Calibration of magnetic field measurement devices in the range of 0.1 mT to 20 mT performed in a Helmholtz coil.  An electromagnet used to generate DC magnetic fields and perform calibrations in the range of 100 mT to 1.6 T. AC magnetic flux density calibration is being offered in the range of 1 µT to 0.5 mT at 50Hz to 20 kHz. Calibration of fluxmeter is also provided in the range of 1 mWb to 10 Wb.

 

m1m2m3m4

DC Helmholtz Coil 

Electromagnet

AC Helmholtz Coil 

Fluxmeter Calibrator 

More Articles ...

                        logo mosti 2          Logo MOF Perbendaharaan Malaysia          sirim           Jabatan Kimia Malaysia         company logo         kpdnkk     JPJ PNG   Logo Agensi Nuklear Malaysia              Logo Jab Alam Sekitar Malaysia res 150x150                                                                                                                                                   APMP logo        aplmf         oiml.png