time 01

NMIM Publications

Calibration of Inductance Standards

Calibration of Inductance Standards
Expired
/ Title

Influence of Adaptor on the Calibration of Inductance Standards

/ Authors

Dewi Mohd Kassim 1, Dan Bee Kim 2, Wan-Seop Kim 3

  1. National Metrology Institute of Malaysia, Lot PT 4803 Bandar Baru Salak Tinggi, 43900 Sepang, Malaysia
  2. Korea Research Institute of Standards and Science, (KRISS) Korea
  3. University of Science and Technology, (UST) Korea.
Image
/ Journal

Journal of Electrical Engineering & Technology, 2018;13(2):911-917

/ Abstract

Influence of the adaptors on the calibration of 100 μH inductance standard was studied as a function of torque, applied when tightening the standard inductor terminal with the adaptor. Two different homemade adaptors of BPO gold-plated brass (BPO-Au) and banana-copper (BN-Cu) were made for the connection between the LCR meter and the inductance standard.

The measured inductance (L) of the standard inductor and the contact resistance (RC) between the adaptor and the standard inductor terminal showed exponential decreases against the torque increase from 25 cN∙m to 150 cN∙m. The measured L and the calculated equivalence series resistance (RS) were dependent on the adaptor type as well as on the RC. The results of the adaptor analysis imply that the BPO-Au adaptor with the lower RC is more suitable for the inductance calibration. The calculated inductance of 99.956 μH corrected by subtraction of the adaptor inductance and the contact resistance  contributions from the measured value using the BPO-Au adaptor agreed well with the certificate (99.948 μH) of the PTB within the measurement uncertainty of 140 μH/H.

Image
National Metrology Institute of Malaysia (NMIM)
Lot PT 4803, Bandar Baru Salak Tinggi
43900 Sepang
Selangor Darul Ehsan

WhatsApp

+6018–2121 833

Corporate Inquiries

irene@sirim.my
+603 - 8778 1605 

Service Inquiries

serviceNMIM@sirim.my
+603 - 8778 1604

Complaint & Feedback

Training Course Request